A 0.6V 2.9µW mixed-signal front-end for ECG monitoring

Established


13
Citations
6
Refs
2012
Year
Search Works
Work Details
  • DOI 10.1109/vlsic.2012.6243792
  • OpenAlex ID W2134612606
  • Type Conference-paper
  • Language EN
  • Published 01 Jun 2012
  • Pages 66–67
  • Source Symposium on VLSI Circuits
  • FWCI 3.44
  • Countries 1
  • Institutions 2
  • Added 24 Jun 2016
  • Updated 16 Oct 2026
MENU
0

A 0.6V 2.9µW mixed-signal front-end for ECG monitoring

Conference-paper 2012 EN Green

This paper presents a mixed-signal ECG front-end that uses aggressive voltage scaling to maximize power-efficiency and facilitate integration with low-voltage DSPs. 50/60Hz interference is canceled using mixed-signal feedback, enabling ultra-low-voltage operation by reducing dynamic range requirements. Analog circuits are optimized for ultra-low-voltage, and a SAR ADC with a dual-DAC architecture…

13

Citations

6

References

3

Authors

3.44

FWCI

Citations by Year
Last 9 years
Topics
Keywords
Authors & Affiliations
3 authors
#AuthorInstitutionsORCID
1 Marcus Yip Massachusetts Institute of Tec… (US) —
2 Jose L. Bohorquez Skulpt (United States) (US)
3 Anantha P. Chandrakasan Massachusetts Institute of Tec… (US)

Data: OpenAlex